Cat. No. 5743 Level 1 - 10 x 1 mL vials
Intended Use
The Cascade Abrazo aPTT assayed controls are intended for use with the Cascade Abrazo analyzer and aPTT test cards to provide a method for quality control of the system. The assayed controls produce clotting times that must be within accepted, standard ranges to indicate that the analyzer and test cards are functioning accurately, and thereby ensure accuracy of the aPTT test card results. These controls can also be used to determine system (analyzer and aPTT test cards) precision.
SUMMARY
Two levels of control plasma are recommended to ensure the accuracy of the aPTT test results. Level 1 provides a sample that will clot within the time range expected for a normal human sample. Level 2 mimics a sample from an individual with a deficiency in intrinsic or common coagulation factors and will exhibit a prolonged clotting time.
REAGENTS
level 1- 1 mL Lyophilized preparation of human plasma and an inert bulking agent. Diluent prepared with deionized water, NaCl, an antimicrobial agent, and an antifoam reagent.
2–8°C (36–46°F) Unreconstituted – until the expiration date on the vial label