Cat. No. 5728 Level 2 - 10 x 1 mL vials
Intended Use The Cascade Abrazo c-ACT-LR assayed controls are intended for use with the Cascade Abrazo analyzer and c-ACT-LR test cards to provide a method for quality control of the system. The assayed controls produce clotting times that must be within accepted, standard ranges to indicate that the analyzer and test cards are functioning accurately, and thereby ensure accuracy of the c-ACT-LR test card results. These controls can also be used to determine system (analyzer and c-ACT-LR test cards) precision.
SUMMARY Two levels of control plasma are recommended to ensure the accuracy of the c-ACT-LR test results. Level 1 provides a sample that will clot within the time range expected for a normal human sample. Level 2 mimics a sample from an individual receiving heparin in low to moderate (up to 3.0 U/mL) concentrations and will exhibit a prolonged clotting time.
REAGENTS Level 2- 1 mL Lyophilized preparation of human plasma and an inert bulking agent. Diluent prepared with deionized water, NaCl, and an antimicrobial agent
2–8°C (36–46°F) Unreconstituted – until the expiration date on the vial label