Gravity air removal in the course of steam pressure is generated in the chamber. Precise proportion with temperature and steam pressure for perfect sterilization of laboratory and other applications
• Elaborated mechanical structure for safety and durability, made of polished stainless steel pressure chamber, door and fabricated stainless steel parts and component for excellent anti-corrosion (Option: STS 316L Chamber)
• Microprocessor PID digital controller operates heating, sterilization temperature and time, steam exhausting automatically with LED process graph indicator and PT-100 temperature sensor
• Heat-proof safety door cover protects user from hot damage or burn, built-in water level sensor prevents the heater from being started in case of water shortage in the chamber and protects autoclave
• Dual safety door lock mechanism prevents starting pro cess in case of unlocked completely and prevents the door from being opened under pressure condition
• With vertical loading type, it is more convenient when you need to sterilize large amount of samples including liquids, solids, etc.
• For safety of the users, the cycle will not start if the door is not locked
• For safety of the users, the door will not open at the end of cycle if the chamber is pressurized
Safety
Over heating temperature protector Water level sensor / shortage alarm
: Automatic over pressure release valve Electric leakage over current circuit breaker Dual door lock safety mechanism